• Global Provider

    Of Instrumentation

  • Technical Design

    Consultation Service

  • Experienced Team

    Industry Professionals

  • Global Provider

    Of Instrumentation

  • Technical Design

    Consultation Service

  • Experienced Team

    Industry Professionals

Model 606-651652 – Portable Tritium in Air Discriminating Monitor

  • Location: State Land, Texas
  • Date: 25/08/2010 - 25/08/2015
  • Estimated Value: $1.000000
  • Field Area: 55.000m2

Portable Tritium in Air Discriminating Monitor

Model: 606-651652

Applications

The 606-651652, Sensetecz Portable Tritium in Air Discriminating Monitor,  combines a 606 unit – Sensetecz Tritium in Air Monitor and a 651652 assembly – Sensetecz Dryer & Catalytic Converter Assembly.

606-651652 has all characteristics of 606. Please refer to the 606 datasheet

Meanwhile 606-651652 has following extra functions:

  • Tritium water and pure elemental tritium readings can be provided

In 606-651652, 651652 is a Dryer & Catalytic Converter Assembly, which is installed between the outlet of the 606-measuring chamber and the inlet of 606-compensation chamber. Using this assembly, the readings of tritium vapor and pure elemental tritium 3H2 can be provided.

Overall, within 0°C to 50°C, 606-651652 can provide reliable readings of active levels of Total Tritium (tritium gas+ tritium vapor) in air, HTO Tritium Vapor,Pure Elemental Tritium 3H2, from 1µci/m³ to 200,000µCi/m³ with full range noble gas compensation and full range gamma compensation. Meanwhile, it can provide gamma activity level readings up to 10R/hr.

Specification – >606-651652 Portable tritium in air discriminating monitor

Performance

  • Sensitivity: Better than 1µci/m³
  • Measuring range: 1µci/m³ to 200,000µCi/m3
  • Accuracy: 1µci/m³ from 1 to 100 µci/m³,  < 1% from 100 to 200,000µCi/m3
  • HTO & 3H2 discrimination / noble gas compensation: Whole measuring range with dryer 651 connected
  • Total tritium, tritium water HTO & elemental 3H2 readings: Extra dryer 651 & catalyst converter tube – model 651652 assembly
  • Flow rate: Up to 1.5L/Min, KNF long life pump
  • Response time: < 10S for more than 90% reading
  • Gamma compensation: Whole measuring range compensation <10µci/m³ at 10mR/hr compensation up to 100mR/hr
  • Gamma measurement: 0.1mR/Hr to 10R/Hr
  • Zero stability / long time drift:         ±2µci/m3 per 30 days
  • Temperature drift: ±2µci/m³ or less, for 0 to 50 °C
  • Noble gas compensation ratio: > 150:1
  • Radon compensation: Via software discrimination and elimination
  • Operation: Real time measuring & alarm control

Structure

  • Ion chamber: 4 chambers, 2 for measuring, 2 for compensation. Measuring volume: 330cc compensation volume: 330cc
  • Ion trap: Yes, built-in for each chamber
  • Ion chamber purging: Yes, with cartridge heater
  • Display: 2.8” TFT LCD, touch screen

Output and communication

  • Linear analog output: 0.5 to 3.3V for 0 to 2800 µCi/m3
  • Logarithm output: 0 to 2.65V for 1 to 200,000 µCi/m3
  • Communication: RS232 default
  • Audio and visual alarm: Low and high alarm, less than 30S
  • Alarm relay output: Contact output for remote control
  • Data logging: Yes, all the readings are time-stamped and logged into 512kbytes EEPROM

Power, operational conditions and dimension:

  • Power consumption: <130mA powered by 12 VDC, external power supply without heater, <350mA powered by two size-D batteries, 25 hours operation, <350mA powered by two size-D NiMH rechargeable batteries capacity 10AmpHr, 15 hrs operation
  • Warm-up time: no need
  • Operational humidity: 95%RH
  • Operational temperature: 0°C to 50°C, the pump’s rated performance might be affected beyond this range
  • Dimension: 4.5”(W)X7”(L)x7”(H) for 606
    & 11”(W)X6”(L)X4.5”(H) for 651652
  • Weight: 7Kgs
© 2016 Sensetecz Engineering | All Rights Reserved