Model 606 – Portable Tritium in Air Monitor

  • Field Area: 55.000m2
  • Location: State Land, Texas
  • Date: 25/08/2010 - 25/08/2015
  • Estimated Value: $1.000000

Portable Tritium Monitor

Model: 606

Applications

606 – Sensetecz Engineering Portable Tritium in Air Monitor is designed to monitor tritium activity levels in in air.

606 uses the Single Body Quadruple Cylindrical Ion Chamber Design (US patent No.: 10725005) to achieve higher grade of gamma and noble gas compensation. An advanced electrometer circuit is used to measure the tritium in air from 1µci/m³ to 200,000µCi/m³ reliably in nuclear power plants and labs, also a Gamma Survey Circuit – a separated GM tube and counting circuit, is built into the unit. This unit monitors the tritium and gamma at the same time and provide the readings for both.

606—Portable Tritium in Air Monitor
The total tritium can be measured and the reading of HTO/HT together will be provided.

606-651—Portable Tritium in Air Noble Gas Compensation Monitor
With an external dryer 651, HTO-tritium water and 3H2 in air can be separated. The readings of the total tritium and HTO tritium water will be provided.

606-651652— Portable Tritium in Air Discriminating Monitor
With an external dryer 651 and a catalyst converter tube 652, the readings of the total tritium, HTO tritium water and pure elemental tritium 3H2 rather than in air will be provided.

Advantages

  • Single body quadruple cylindrical ion chamber design US Patent Number: 10725005One stainless block with 4 machined hollow cylinders to get the best volume uniformity and symmetrical orientation, therefore higher grade of gamma and noble gas compensation is achieved, also reduces the leakage and simplifies the construction.
  • Full range noble gas compensation and no range switching required

Ultra precision/stable electrometer circuit with 24 bits ADC convertor to achieve high sensitivity – better than 1µci/m³ and high measuring range – 1µci/m³ to 200,000µCi/m3 without range switching. When combined with Sensetecz external drier 651, noble gas compensation within full measuring range can be achieved. This fixes the saturation problem in high noble gas compensation.

  • Drier Monitoring by humidity sensor

A precision humidity sensor is installed at the inlet of compensation chambers. A drier fault will be generated when humidity of compensation chamber is higher than the setting. The fault will be sent out via RS232 to the position of Flow Alarm to ensure safe operation of noble gas compensation. 

Performance enhancement  

  • Good Stability, Ensured ±2uCi/m3 or less for 30 days and for temperature from  0°C to 50°C. No Negative reading shown under low temperature condition.
  • Full measuring range noble gas compensation, no range switching for full measuring range
  • Dryer status is monitored to ensure safe noble gas compensation (A dryer fault will triger the existing flow alarm contact open and send out the flow alarm via RS232).
  • No chamber internal leakage due to the single body ion chamber construction .
  • Built-in ion trap with high electrical field to reduce charge particle interference.
  • Improved operation and maintenance ability. Simplified the construction and less mechanical parts used.
  • Standard Batteries are used:
  • Four of A23 12V batteries are used for ion chamber bias and can last 5 years
  • Four of AA 1.5V batteries are used for powering electrometer. They are rated 2.5Ahr and can last 2 years
  • Two D batteries (re-chargeable or non-rechargeable) are used for pump and display
  • Fast response, less than 10 seconds will reach 90% of reading
  • All alpha radiation, including radon daughter compensation are accomplished by electronic circuit and software filter
  • Environmental gamma measuring is accomplished by a GM tube with the corresponding circuit
  • Ion chamber decontamination by a cartridge heater
  • LEDs and LCD background color changes indicate the alarm and warning states
  • On-Line NiMH battery smart charger circuit with rechargeable battery recognition is used to avoid charging the non-rechargeable Meanwhile the batteries are thermo-state switch protected.
  • Low power consumption free scale MKL25 microcontroller is used for the control/display
  • Data logging – all the readings are Time-stamped and logged into 512kbytes EEPROM

Specification – 606 portable tritium in air monitor

Performance

  • Sensitivity: Better than 1µci/m³
  • Measuring Range: 1µci/m³ to 200,000µCi/m3
  • Accuracy: 1µci/m³ from 1 to 100 µci/m³, < 1% from 100 to 200,000µCi/m³
  • HTO & 3H2 discrimination / noble gas Compensation: Whole measuring range with dryer 651 connected
  • Total tritium, Tritium water HTO & elemental 3H2 readings:  Extra dryer 651 & catalyst converter tube 652
  • Flow rate:  Up to 1.5L/Min, KNF long life pump
  • Response time: < 10S for more than 90% reading
  • Gamma compensation: Whole range compensation <10µci/m³ at 10mR/hr compensation up to 100mR/hr
  • Gamma measurement: 0.1mR/Hr to 10R/Hr
  • Zero stability / long time drift: ±1µci/m3 per 30 days
  • Temperature drift: 2µci/m³ or less, for 0 to 40 °C
  • Noble gas compensation ratio: > 150:1
  • Radon Compensation: Via software discrimination and elimination
  • Operation: Real time measuring and alarm control

Structure:

  • Ion Chamber: 4 chambers, 2 for measuring, 2 for compensation. measuring volume: 330cc compensation volume: 330cc
  • Ion trap:  Yes, built-in for each chamber
  • Ion Chamber purging: Yes, with cartridge heater
  • Display: 2.8” TFT LCD, touch screen

Output and communication:

  • Linear analog output: 0.5 to 3.3V for 0 to 2800 µCi/m3
  • Log output: 0 to 2.65V for 1 to 200,000 µCi/m3
  • Communication: RS232 default
  • Audio and visual alarm:  Low and high alarm, less than 30S
  • Alarm relay output: Contact output for remote control
  • Data logging: Yes, all the readings are time-stamped and logged into 512kbytes EEPROM

Power, operational conditions and dimension:

  • Power consumption:  < 130mA at 12 VDC input, external power supply without heater,<350mA powered by two size-D batteries, 25 hours operation,<350mA powered by two size-D NiMH rechargeable batteries capacity 10AmpHr, 15 hrs operation
  • Warm-up time:  no need
  • Operational humidity: 95%RH
  • Operational temperature: 0 to 50 °C
  • Dimension:  4.5(W)X7(L)x7(H) including handle
  • Weight:  5Kgs

 