Wall Mounted Tritium in Room Air Monitor
Model 607C-0.8U0.8CI
Applications
607C-0.8U0.8CI – Sensetecz Engineering Tritium in Air Monitor (hereinafter referred as 607C) is a 19” rack mount type instrument. It is customized from default 607 model.
607C-0.8U0.8CI has all characteristics of default 607 except its customized sensitivity and measuring range. Please refer to the model 607 datasheet for its default characteristics.
The customizations of 607C-0.8U0.8CI are as below:
- Sensitivity: 0.8µCi/m3
- Measuring range: 0.8µCi/m3 to 0.8Ci/m3.
- 8µCi/m3 to 0.8Ci/m3 is defined as the full measuring range for all characteristic for 607C-0.8U0.8CI, i.e. Full measuring range noble gas compensation, full measuring range gamma compensation, etc.
Specification
Performance:
- Sensitivity: Better than 0.8µci/m³
- Accuracy: < 1% from 100 µci/m³ to 0.8Ci/m3 1µci/m³ from 1 to 100 µci/m³
- Measuring range: 0.8µci/m³ to 0.8Ci/m3
- Compensation range: Full measuring range, 0.8µci/m³ to 0.8Ci/m3
- Flow Rate: 1 to 10L/min, KNF positive displacement diaphragm pump, brushless, 10000hrs and up
- Response time: < 10S
- Gamma measurement: 0.1mR/Hour to 10R/Hour
- Zero Stability: Better than 2µci/m³ from 0 to 50 °C
- Stability with long term: <±3% per 30 days
- Gamma compensation: <10µci/m³ at 10mR; <10% within 20mR/HR
- Tritium Discrimination: Elemental tritium, HTO & T2O discrimination via 651 Dryer, 652 catalyst converter and software computation
- Noble Gas Cancellation: Full Measuring range, Passive Silica gel drier is placed on the front panel, flowing air samples between measuring chambers and compensation chambers
- Radon Compensation: Software discrimination and elimination
- Background Cancellation: Noble gas compensation ratio: > 150:1. Gamma compensation: <10µci/m³ at 10mR/Hr, <10% within 20mR/HR
Structure:
- Ion Chambers: 4 ion chambers, 2 for measuring and 2 for compensation, 500mL for each chamber
- Ion Chamber Volume: Default, 1L: measuring, 1L: compensation
- Ion Trap: Yes, built-in for each chamber
- Gas Ports: 3/8” push to collect
- Air Hose: Default, 3 feet for inlet and outlet
- Gas Filter: Built-in, replaceable
- Ion Chamber Purging: Yes, via cartridge heater for ion chamber decontamination and condensation prevention
- Control Panel Display: 7 Inch TFT LCD touch screen for numerical and graphical model and operation; LED for digital data and rate bar graph
- Digits Display: 0.56” LED display both for tritium and Gamma
Output and communication:
- Communication: RS232, RS485 and ethernet
- Analog Output: 4-20mA and 0-3Vdc
- Audio and Visual Alarm:Low and high alarm within 30 seconds
- Data Logging: Yes, frequency varies from 1s to 1hr, user settable
Power, operational conditions and dimension:
- Power Supply: 220/115VAC @ 50/60HZ
- Power Consumption: < 1A
- Warm-up time: no need
- Operational Humidity: 95%RH
- Operational Temperature: 0 to 50 °C
- Dimension: 5U, W19”x H8.72” x D10.12”
- Weight: 20Kgs